Maintenance
CR 35-X D
IGITIZER
4
Hold the wire with two fingers
near the housing
and push it carefully towards the
housing.
5
Repeat this sequence until the head of the wire has reached the housing.
6
Then draw out the wire to its outmost position and repeat cleaning once
again.
7
Fix the wire to the housing again.
Note: Misuse of the bowden wire leads to bending. This causes a complicated