Using a moving grid
CAUTION:
The connected examination table or the bucky must be in
compliance with IEC 60601-1 standard and must be connected
to the equipment according to 93/42/CE medical devices
directive and following amendments.
1.
Connect the examination table or the bucky.
Figure 27: Connector for moving grid on a system with rotating column
Figure 28: Connector for moving grid on a system with fixed column
2.
Activate the moving grid.
Figure 29: Activate moving grid
The indicator lights up when the signal for the moving grid is enabled.
3.
Prepare the exposure.
4.
Execute the exposure.
If exposure is blocked, check the error messages and the connection to the
bucky.