![AGFA 6012/001 Скачать руководство пользователя страница 111](http://html.mh-extra.com/html/agfa/6012-001/6012-001_user-manual_2866620111.webp)
DAP meter
Model
Diamentor CI-P
Type
Device for the dose area product meas-
urement in X-ray diagnostics according
to IEC 60580 standard.
Principle of measurement
Radiation measure with ionization
chamber
Measured quantity
Dose area product
Digital resolution
0.01
μ
Gym
2
Max linearity error
< 2.5%
Nominal range of dose-area prod-
uct rate
(0.01 - 2500)
μ
Gym
2
/s
Equivalent filtration of the cham-
ber @75kV
0.3 mm Al
Max measurement field
118 mm x 118 mm
Dimensions (W x D x H)
152 mm x 234 mm x 23 mm
Weight
455 g