INVERTER KV ERROR
During the exposure,
the kV value has de-
creased under 75% or
increased over 110% of
the requested value.
Unbalance in the high
voltage circuit during
the exposure
Press the reset button
and repeat the expo-
sure.
INV. OVERLOAD
Power to the inverter is
out of range.
Press the reset button to
continue.
INVERTER ERROR
IGBT drivers error.
Press the reset button to
continue.
TUBE CALIB. ERR.
Error of X-ray tube cali-
bration.
Call Service.
XR HANDSWITCH ERR. Faulty exposure button. Check the integrity of
the exposure button,
switch off the device
and switch it on, try
again, if the error per-
sists call Service.