![AGFA 6012/001 Скачать руководство пользователя страница 78](http://html.mh-extra.com/html/agfa/6012-001/6012-001_user-manual_2866620078.webp)
Using the remote control exposure button
The remote control is optional.
Ensure that there is no other equipment with remote control nearby.
Keep as fas as possible from the radiation source. The remote control operates
at a distance of maximum 11 m from the device at any angle and within sight,
through glass or leaded glass. The remote control doesn't operate through
doors or walls.
1.
Point the remote control towards the remote control receiver on the
device.
2.
Press and release the exposure button on the remote control.
The X-ray tube is prepared for performing an exposure.
3.
Press the exposure button withing 15 seconds and keep it pressed until the
exposure has ended.
When the X-ray exposure is finished successfully, three beeps are heard.
WARNING:
During exposure ionizing radiation is emitted by the X-ray
system. To indicate the presence of ionizing radiation, the
radiation indicator on the control console lights up.
4.
Return the remote control to its cradle.
The remote control produces a sound signal if it is not returned to its
cradle after three minutes.
Related Links
on page 28
78
| DR 100e (Digital Radiography) | Basic Workflow
0365B EN 20181015 0853