Harmonic emis-
sion in accord-
ance with IEC
61000-3-2
frequency communication services. The user
might need to take mitigation measures, such
as relocating or re-orienting the equipment.
Class A
Voltage fluctua-
tions / flickering
in accordance
with IEC
61000-3-3
Fulfilled
The device is used in a professional healthcare / radiological environment.
Environmental conditions are stated in the user manual.
This device was tested for a professional healthcare environment as described
above. Nevertheless the HF-emission and immunity can be influenced by
connected data cables depending on length and the manner of installation.
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
Resistance to Jam-
ming Test
Test level of profes-
sional medical
equipment and ba-
sic EMC standards
Electromagnetic Environ-
ment Guidelines
Discharge of static
electricity in accord-
ance with IEC
610004-2
+ 8 kV contact dis-
charge
+ 15 kV air dis-
charge
Floors should consist of
wood, concrete or ceramic
tiles. The relative humidity
must be at least 30%, if the
floor is made of synthetic
material.
Fast transient electri-
cal disturbance varia-
bles / bursts in accord-
ance with IEC
610004-4
+ 2 kV mains
+ 1 kV data lines
The quality of the voltage
supplied should correspond
to a typical commercial or
clinical environment.
Impulse voltages
(surges) in accordance
with IEC 61000-4-5
+ 1 kV line-line volt-
age
+ 2 kV line-ground
voltage
The quality of the voltage
supplied should correspond
to that of a typical commer-
cial or clinical environment.
Voltage breakdown,
short term interrup-
tions and variations in
the voltage supplied in
• 0% U
r
for ½ peri-
od
• 0% U
r
for 1 peri-
od
• 70% U
r
(30%
breakdown of
The quality of the voltage
supply should correspond to
that of a typical commercial
or clinical environment.
118