![AGFA 6012/001 Скачать руководство пользователя страница 102](http://html.mh-extra.com/html/agfa/6012-001/6012-001_user-manual_2866620102.webp)
DR 100e
Electrical connection
Rated power supply
115 ÷ 230Vac ±10%, 50/60 Hz ± 5
Hz, 10A, standard monophase with
ground conductor.
Automatic setting of the equipment
according to mains voltage.
Automatic line compensation.
Standard outlet
16 A @ 230Vac
Line resistance
<1
Ω
@ 115/230Vac
Power supply cable
8 m
Infrared remote control
2 x 1.5 V alkaline battery, 2500mAh
Battery life > 25000 exposures
Radiological data
Nominal power (IEC 60601-1)
32 kW @ 100 kV, 320 mA, 100 ms
kV values
40 - 125 kV at steps of 1 kV
kV accuracy
±5% (IEC 60601-2-54)
mA values @115/230Vac
50 - 400 mA
mA accuracy @115/230Vac
±10% (IEC 60601-2-54)
mAs values @115/230Vac
0.1 - 220 mAs
mAs accuracy
±10% (IEC 60601-2-54)
Exposure times @115/230Vac
0.001 - 2.2 s according to mAs
Time accuracy
±10% (IEC 60601-2-54)
Max mAs
40 kV
max. 220 mAs
(the optional second capacitor bank is
required)
41 - 45 kV
max. 200 mAs
46 - 52 kV
max. 180 mAs
53 - 62 kV
max. 160 mAs
63 - 72 kV
max. 140 mAs
102
| DR 100e (Digital Radiography) | Product Information
0365B EN 20181015 0853