New devices are delivered without preset directories and files:
13 Operation
13.1
Notice concerning measurement of SO2 and NO2 (option)
SO
2
and NO
2
gases have a high solubility in water. For measure-
ment of SO
2
and NO
2
concentrations it is therefore necessary to re-
move the condensate residues form the gas filtration and drying sys-
tem. These residues can absorb SO
2
and NO
2
which could cause
measurement deviations.
Furthermore, when carrying out SO
2
and NO
2
relevant measure-
ments no additional desiccant should be used. Even when it is dry
this filter material can absorb significant parts of the SO
2
and NO
2
content.
13.2
Battery/line voltage operation (only for the measurement
module)
•
Battery mode: Battery operation: Up to 36 hours of continuous
operation with the display backlight switched on.
•
Charging: Intelligent charging by means of an integrated
charger management system.
•
Battery operation is limited to the measurement module MAX-
ILYZER NG Plus. The Peltier gas cooler (with the heating con-
troller) can only be operated via mains.