![AFRISO MAXILYZER NG Plus Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/afriso/maxilyzer-ng-plus/maxilyzer-ng-plus_instruction-manual_2864828021.webp)
Transportation and storage
MAXILYZER NG Plus
21
5
Transportation and storage
CAUTION
Damage to the device due to improper transportation.
Do not throw or drop the device.
CAUTION
Damage to the device due to improper storage.
Protect the device against shock when storing it.
Store device in a clean and dry environment.
Store device only within the admissible temperature range.
Store device away from paint, solvent and glue.
6
Commissioning
Commissioning the Peltier cooler
1. Connect MAXILYZER NG Plus to mains.
2. Actuate the "ON/OFF" switch (Peltier cooler and hose heating
controller).
The connected system components are heated up to operating
temperature. This may take up to 15 minutes.
When the green status LEDs light up, the device is ready for
operation.
3. You can start measurements once all system components are
ready for operation.
If gas cooling is not required, it is possible to operate MAXILYZER
NG Plus with the gas cooler switched off.