Commissioning
MAXILYZER NG Plus
23
6.2
Mounting and installation
MAXILYZER NG Plus is a portable device. To ensure proper
separation and discharge of the condensate in the cooler, MAX-
ILYZER NG Plus must be placed on an even, horizontal sur-
face.
Use MAXILYZER NG Plus only away from heat sources and
when free ventilation is ensured to avoid heat accumulation in
the device. The fan openings must not be covered.
CAUTION
Damage to the device due to excessively high or low ambient
temperatures.
Operate MAXILYZER NG Plus only in the admissible ambient
temperature range from +5°C to +40°C.
6.3
Electrical connections
The correct mains voltage is available.
MAXILYZER NG Plus is shipped with a low temperature cable (1.5 m
long).
Connect the low temperature cable to the main socket on the
front plate.
Heating hose and sampling probe
The electrical power of the heating hose and the sampling probe
does not exceed 600 W each.
Use the round connector with the designation "Heating Hose" to
supply the gas sampling probe and the heating hose.
The sampling probe can also be connected directly to MAXILYZER
NG Plus without the heating hose.
6.4
Hose connections
CAUTION
Damage to the device due to improper use.
If you measure corrosive gases, make sure there are no com-
ponents damaging the gas transport parts.
Unheated gas sampling lines should have a gradient to the cooler in
order to avoid condensate accumulations in the gas supply line.
If you use a heated gas line, it must be supported in such a way as
to prevent bends in the gas supply line.