Specifications
20
MAXILYZER NG Plus
Parameter
Value
Sampling probe
(Standard) type
V2A 1.4571 (300 mm)
4.1
Calculation formulae (extract)
Calculation of the CO
2
value
CO
2
= CO
2 max
* (1 -
21
O
2
) in %
CO
2max
Max. CO
2
value (fuel-specific) in % Volume
O
2
Measured oxygen content in % Volume
21
Measured oxygen content in % Volume
Calculation of the waste gas loss
qA = (T.Gas - T.Air) * (
2
2
O
21
A
−
+ B) in %
T.Gas Waste/flue gas temperature in °F or °C
T.Air
Combustion/ambient temperature in °F or °C
A2, B
Fuel-specific factors
Calculation of the excess air value Lambda
Lambda =
2
max
2
CO
CO
=
2
O
21
21
−
Calculation of the combustion efficiency value (Eta)
Eta = 100 - qA in %
Calculation of CO 0% (undiluted)
CO
unv.
= CO * Lambda
CO
unv.
Content of carbon monoxide, undiluted
CO
Reading for CO
4.2
Approvals, tests and conformities
Maxilyzer NG Plus is approved as per 1
st
German Federal Immission
Act (Bundesimmissionsschutzverordnung) and EN 50379-2. It is
TÜV-tested and complies with the applicable directives 2004/108/EC
(EMC) and 2006/95/EC (low voltage).