Menu “Measure”
MAXILYZER NG Plus
43
Function “CO protect”
Every device is equipped with a second pump (CO-flushing-pump) in
order to protect the quite sensitive CO-sensors from CO-overload.
The CO-flushing-pump can either be started manually or it switches
on automatically when necessary, i.e. when the admitted CO-range
is exceeded.
Register
Key
Function
Switch CO-flushing-pump on/off.
If the CO-flushing-pump starts automatically
due to an excess concentration of CO it can’t
be switched off manually until the high CO-
concentration is no danger for the CO-Sensor
anymore. If the CO-concentration has rea-
ched the lower range again the CO-flushing-
pump will shut off.
When the CO-flushing-pump is activated a scored out CO-symbol
appears in the status line:
If the device is equipped with two CO-sensors the result of the higher
range sensor will be displayed when the lower range sensor is flus-
hed.
The active CO-flushing-pump doesn’t influence any other sensors
within the device.
Function “Print”
Register
Key
Function
Print measured data.
The printer in use can be selected from the configuration menu, refer
to chapter 11.3, page 57. The rate of printing depends mostly on the
type of printer selected.
Because of the modern multi-tasking-operating the device can be
used without restrictions during the printing procedure. Printing takes
place simultaneously to the other procedures in order to avoid de-
lays.