Ensemble LAB Hardware Manual
Installation and Configuration
2.5.5. High-Speed User Inputs 4-5
The high speed inputs 4-5 are typically used as a sample signal for data collection.
Table 2-23:
High-Speed Input Specifications
Specification
Value
Input Voltage
5V or 24 V
Input Current
10 mA
Input Device
HCPL-0630
Delay
50 nsec
Table 2-24:
Port 0 High Speed Digital Input Connector Pin Assignment (J106)
Pin#
Description
In/Out/Bi
3
High-Speed Input 4 + user interrupt
Input
4
High-Speed Input 4 - user interrupt
Input
5
High-Speed Input 5 + user interrupt
Input
6
High-Speed Input 5 - user interrupt
Input
J106-3
J106-4
J106-5
J106-6
HCPL-0630
3
3
0
3
3
0
CURRENT LIMITER
+5V
+5V
OPTO4+
OPTO4-
OPTO5+
OPTO5-
CURRENT LIMITER
Figure 2-28:
High Speed User Inputs (J106)
www.aerotech.com
Chapter 2