Ensemble LAB Hardware Manual
Installation and Configuration
2.5.4. Opto-Isolated Inputs 0-3
The digital inputs are opto-isolated and may be connected to current sourcing or current sinking devices, as
shown in
and
. These inputs are designed to connect to other ground-referenced
circuits and are not intended for high-voltage isolation.
The opto-isolator's common connections can be directly connected to the drive's power supply; however,
doing so will effectively defeat the isolation and will reduce noise immunity.
Table 2-21:
Opto-Isolated Input Specifications
Input Voltage
Approximate Input Current
Turn On Time
Turn Off Time
+5 V
1 mA
200 usec
2000 usec
+24 V
6 mA
4 usec
1500 usec
Table 2-22:
Port 0 Digital Input Connector Pin Assignment (J106)
Pin#
Description
In/Out/Bi
17
Opto-Isolated Input 0 / CCW EOT Input
(1)
Input
18
Opto-Isolated Input 1 / CW EOT Input
(1)
Input
24
Input Common
-
25
Opto-Isolated Input 2 / Home Input
(1)
Input
26
Opto-Isolated Input 3
Input
(1) Software configured option
www.aerotech.com