Sweep Type = LIN for all tests
Where applicable, frequency values should be within 2% of values selected
For the Square waveform, slew rate should be within 3 V/ms for each test
Waveform
Start/End Offset
(V)
Frequency
(kHz)
Measured Frequency
(kHz)
Low Limit
(V)
Measured
(V)
High Limit
(V)
DC
0
NA
NA
0.0
0.0
100
NA
NA
4.9
5.1
200
NA
NA
9.9
10.1
Ripple
Start/End
Amplitude, Vp
Frequency
(kHz)
Measured Frequency
(kHz)
Low Limit
(Vp)
Measured
(V)
High Limit
(Vp)
100
1
4.8
5.3
10
4.8
5.3
100
4.8
5.3
300
3.5
3.6
200
1
9.5
10.5
10
9.5
10.5
20
9.5
10.5
Sawtooth
Start/End
Amplitude, Vp
Frequency
(kHz)
Measured Frequency
(kHz)
Low Limit
(V)
Measured
(V)
High Limit
(V)
100
1
4.8
5.3
10
4.8
5.3
20
4.8
5.3
200
1
9.5
10.5
10
9.5
10.5
20
9.5
10.5
Phase*
Start/End
Vp = 20 V, f = 1
kHz
Phase
Offset (°)
Measured Frequency
(kHz)
Low Time
(msec)
Measured
Time
(msec)
High Time
(msec)
90
NA
735
765
180
NA
490
510
270
NA
245
255
Table 1
Содержание DSR 100-15
Страница 8: ...7 Press the Save button to save the resulting calibration numbers and close the Settings window ...
Страница 9: ......
Страница 18: ...useful for testing a sequence before generating output ...
Страница 65: ... See procedure above If you have any questions please contact AE Techron Technical Support at 574 295 9495 ...