System Calibration
IMPORTANT:
Before operating the DSR 100-15, the Calibration Test should be run to determine the correct system
settings for the DSR 100-15.
The DSR 100-15's default System Gain setting is 20, and the default DC Offset setting is 0, but these settings should be
verified and adjusted, if needed, to reflect the measured gain of your DSR 100-15 system at the Test Supply outputs. If
you are measuring the output of the DSR 100-15 without amplification (from the Controller's front-panel BNC signal
output), a System Gain of 1 can be used for DSR 100-15 calibration (for an expected output of
10Vp/20Vpp).
Complete the following steps to run the calibration test and adjust the DSR 100-15's System Gain and DC Offset settings:
1. Select the Settings button and then open the System Calibration tab.
2. Make sure an estimated gain is entered in the Amplifier Gain input box. A default gain of 20 will be pre-entered in
this box.
3. Adjust the Calibration Test Voltage, if desired. A default test signal of 1 Vp will be pre-entered in this box. Note: the
default DC offset is 0 and cannot be changed.