![Advanced Instruments GPR-IN190 Скачать руководство пользователя страница 34](http://html.mh-extra.com/html/advanced-instruments/gpr-in190/gpr-in190_owners-manual_2856039034.webp)
Advanced Instruments Inc.
Liquid covering sensing area
Improper sensor selection
Presence of interference gases
Unauthorized maintenance
Sensor nearing end of life
Wipe with alcohol and lint free towel or flow sample
or zero gas for 2-3 hours to flush
Replace GPR/PSR sensor with XLT sensor when CO
2
or acid gases are present. Consult factory.
Replace sensor and install scrubber
Consult factory.
Replace sensor
Erratic O
2
reading
or
Negative O
2
reading
or
No O
2
reading
accompanied by
electrolyte leakage
Pressurizing the sensor by flowing gas to the
sensor with the vent restricted or SHUT OFF
valve closed and suddenly removing the
restriction draws a vacuum on the sensor
or
partially opening the valves upstream of the
analyzer when using a pump downstream of
the analyzer to draw sample from a process at
atmospheric pressure or a slight vacuum.
Placing a vacuum on the sensor in excess 4”
of water column is strongly discouraged.
A premature adjustment of the ZERO OFFSET
potentiometer is a common problem
Zero the analyzer. If not successful replace the
sensor
Avoid drawing a vacuum on the sensor, a
pressurized sensor may not leak but still produce
negative readings.
From MAIN MENU select DEFAULT ZERO
33