GAS CONCENTRATION
85.0 PPM
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
>>>
87.4 PPM
SPAN CALIBRATION
WAIT FOR STEADY RDG
ENTER TO CALIBRATE
MENU TO ABORT
By using the UP or DOWN arrow keys, enter the appropriate digit where the cursor is blinking
Press the ENTER key to advance the underline cursor right or press the MENU key to advance the underline cursor
left to reach to the desired digit of the gas value.
Repeat until the complete span value has been entered.
In the example above, a span value of 85.0 PPM has been entered.
After the span value has been entered, the analyzer will display the actual oxygen reading and prompt to press the
ENTER key to accept SPAN CALIBRATION or MENU to escape.
Caution: Allow the analyzer reading to stabilized before accepting calibration.
After successful calibration, the analyzer will display a message “Passed Calibration” and return to the Sample mode.
NOTE: The analyzer is allowed to accept calibration when O2 reading is within the acceptable value. If the O2
reading is outside of this limit, by pressing ENTER to accept calibration will result in “Failed Calibration” and return to
the Sample mode without completing Span calibration. After pressing ENTER either of the following two messages
will be displayed and the analyzer will return to SAMPLE mode.
PASSED
CALIBRATION
OR
FAILED
CALIBRATION
Default
Span
The software will set the SPAN adjustment based on the average output of the oxygen at a specific oxygen
concentration and erase any previous span calibration data. For example, with factory default settings, when a span
gas is introduced, the micro-processor will display oxygen reading 30-50% of the span gas value, indicating
that the sensor output is within the specified limits. This feature allows the user to check the sensor’s signal output at
a specified oxygen concentration without removing it from the sensor housing.
Access the MAIN MENU by pressing the MENU key.
Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
Press the ENTER key to select the highlighted menu option.
The following display appears:
MAIN MENU
SELECT RANGE
CALIBRATION
VIEW HISTORY
SYSTEM OPTIONS
>>>
CALIBRATION
SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN
DEFAULT ZERO
OUTPUT CALIBRATE
OUTPUT SIMULATE
Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT SPAN.
Press the ENTER key to select the highlighted menu option.
25