![Advanced Instruments GPR-IN190 Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/advanced-instruments/gpr-in190/gpr-in190_owners-manual_2856039028.webp)
Advanced Instruments Inc.
Analog
Output check-Output Calibrate
In certain cases, the full scale analog may not match with full scale display. This feature allows the user to adjust the
electronics so that the full scale display matches with full scale analog signal output. To calibrate full scale signal
output
Access the MAIN MENU by pressing the MENU key.
Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION and then select OUTPUT
SIMULATE.
Press the ENTER key to select the highlighted menu option.
The following displays appear:
MAIN MENU
SELECT RANGE
CALIBRATION
VIEW HISTORY
SYSTEM OPTIONS
>>>
CALIBRATION
SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN
DEFAULT ZERO
OUTPUT CALIBRATEI
OUTPUT SIMULATE
OUTPUT SPAN
20.0 mA ADJUST
PRESS UP OR DOWN TO ADJUST
OUTPUT ENTER/MENU TO RETURN
Pressing UP or DOWN key will increase or decrease the full scale output signal each time. Check the output on the
external recording device or voltmeter/ammeter. Repeat this step until the out equals the full scale analog signal
expected, for example 20 mA in the present case. After OUTPUT CALIBRATION is complete, press ENTER/MENU key
to return to SAMPLE mode.
Sampling a Gas
GPR-1500 Oxygen Analyzer requires a positive pressure to flow the sample gas across the sensor to measure the
oxygen concentration in a sample gas. If a positive sample pressure is not available, install a high quality external
sample pump to push the sample through the analyzer; see the option of using a sample pump as described above.
Procedure
Following calibration, the analyzer will return to the SAMPLE mode and ready for sampling the gas.
Select the desired sampling mode - auto or manual – as described above.
Use a suitable tubing to transport the sample gas to the analyzer
The main consideration is to eliminate any air leaks which can affect oxygen measurements.
For sample gases under positive pressure, the user must provide a means of controlling the inlet pressure between
5-30 psig.
For sample gases under atmospheric or slightly negative pressure, an external pump is necessary to push the sample
through the sensor housing. Generally, when using a low voltage DC pump, no pressure regulation is necessary but a
flow control device is recommended; a flow meter upstream of analyzer is recommended to ensure that the sample
flow is adequate and steady.
Assure the sample is adequately vented for optimum response and recovery – and safety.
27