Technical Specifications *
Accuracy:
< 2% of FS range under constant conditions
Analysis:
0-10, 0-100, 0-1000 PPM, 0-1%, 0-25% (CAL) FS
Auto-ranging or manual lock on a single range
Application:
Oxygen analysis in inert, helium, hydrogen, mixed and
acid (CO
2
) gas streams
Approvals: CE
Area Classification: General purpose
Alarms:
Two user configurable alarms: magnetic coil relays rated
3A at 100 VAC, field programmable alarm time delays,
alarm bypass for CAL and system fail alarm
Calibration:
1 month interval using certified span gas or ambient air
Compensation:
Barometric pressure and temperature
Connections:
1/8" compression tube fittings
Controls:
Water resistant keypad; menu driven range selection,
calibration, alarm and system functions
Display:
Graphical LCD 5” x 2.75”; resolution .01 PPM; displays
real time ambient temperature and pressure
Enclosure:
Painted aluminum 6” x 4” x 4” panel mount
Flow:
1-2 SCFH recommended
Linearity:
> .995 over all ranges
Pressure:
Inlet - regulate to 5-30 psig to deliver 1-2 SCFH flow;
vent - atmospheric
Power:
Universal; specify 100/120/220/240 VAC or 12-28 VDC
Range ID:
4-20 mA or relay contacts
Response Time:
90% of final FS reading < 10 seconds
Sample System:
Sample Flow-through sensor housing
Sensitivity:
< 0.5% of FS range
Sensor Model:
GPR-12-333-M for non-acid (CO2) gas streams;
XLT-12-333-M for gases containing > 0.5% CO2
Sensor Life:
24 months in < 1000 PPM O2 at 25ºC and 1 atm
Signal Output:
4-20mA
Temp. Range:
-10ºC to 45ºC (GPR), -20ºC to 45ºC (XLT sensor)
Warranty:
12 months analyzer; 12 months sensor
Wetted Parts:
Stainless steel
Optional Equipment
2-part SS sensor housing, Sample conditioning accessories - contact factory
* Specification subject to change without notice.
Recovery Time:
60 seconds in air to < 10 PPM in < 1 hr on N2 purge
Advanced Sensor Technology
Unmatched Expected Life & Warranty
Unmatched Recovery from Air to 10 PPM
Extended Operating Range to –10
⁰
C
Excellent Compatibility with 0-100% CO
2
Sensitivity < 0.5% FS Range
Fast Response, No Maintenance
100-240 VAC or 12-28 VDC Power
2 Field Selectable Alarm Set-points
Power /System Failure Alarms
5 Ranges Standard
4-20 mA
Optional Range ID 4-20 mA or Relay
Contacts
Stainless Steel Wetted Parts
IND-190
PPM Oxygen Analyzer
Shown with optional 2-part SS sensor housing
Panel Mount Configuration for OEM Applications
ISO 9001:2008 Certified
INTERTEK Certificate No. 485