GB 01725 - Edition 05 - Nov 12
1/5
A 400
Technical characteristics of the pump
Characteristics
UNIT
ATH 2300 M
ATH 2300 MT
Flange (inlet)
ISO-K
DN 200
DN 250
DN 200
DN 250
Flange (exhaust)
ISO KF
DN 40
Pumping speed*
Ar
l/s
1 450
2 000
1 450
2 000
H2
l/s
1 100
1 100
1 100
1 100
He
l/s
1 650
1 800
1 650
1 800
N2
l/s
1 550
2 150
1 550
2 150
Compression rate
Ar
> 1 · 10
8
H2
> 2 · 10
3
He
> 3 · 10
4
N2
> 1 · 10
8
Gas throughput
(1) (2)
Ar
sccm
1 500
500
H2
sccm
> 4 000
> 1 000
He
sccm
> 4 000
> 1 000
N2
sccm
4 000
1 000
Fore vacuum max.
(3)
Ar
hPa
3.7
3.5
H2
hPa
0,75
He
hPa
2
N2
hPa
2.9
2.8
Inlet vacuum max.
(2)
Ar
hPa
0.025
0.006
N2
hPa
0.25
0.015
Ultimate pressure
(4)
hPa
< 6 · 10
-9
Nominal rotation speed
tr/mn (Hz)
31 000 (517)
Standby speed
tr/mn (Hz)
from 15 000 (250) to nominal speed 31 000 (517)
Sound pressure level
db(A)
< 47
Pump protection index
IP 42
Controller power supply
V
200-240 (50/60 (Hz)
Maximum leakage current
mA
< 12
Max. start-up supply **
W
< 1 000
< 1 250
Power consumption at ultimate pressure
W
< 130
< 380
Max. apparent start-up power **
VA
< 1 700
< 1 900
Mounting orientation
any
Maximum heating temperature
°C
-
75
Heating power of the heating band (the
same input voltage as the controller)
W
-
250 (nominal value)
N2 purge flange
ISO KF
DN16
Integral leak rate
(7)
hPa l/s
< 5·10
-8
* Without inlet screen.
** Value could be depend on controller type.
(1)
At nominal rotational speed, water temperature = 25 °C (M = cold ; MT = 65 °C).
(2)
With a low exhaust pressure, depends on external conditions (water temperature, water flow, ambient temperature).
(not applicable for air cooling version).
(3)
With ratio inlet pressure/exhaust pressure >100, max. throughput reduced.
(4)
ISO/F flange, better with CFF flange.
adixen Vacuum Products - Operating instructions - ATH 2300 M/MT