2/2
B 340
Water cooling connection
GB 01725 - Edition 04 - Nov 12
For ATH 2300 MT
model
- Provide a water inlet
pipe and a tap to adjust the flow
rate.
- Connect the water valve to the
water inlet line using a flexible tube
following the assembly diagram:
- Connect the other nipple to the
draining circuit.
Wate IN
Water OUT
Valve connection
This pump is equipped with a heating band and a water valve. To guarantee
the performance of the MT pump on difficult processes, make sure to:
- Connect the heating device, set the heating temperature via the controller
(see Operating Instructions of the controller).
- Connect the water valve, check the efficiency of the cooling system and
the absence of leaks. If stopping the cooling circuit creates a significant risk,
install a «water sensor» security on the circuit.
Heater band connection in the Operating instructions of
controller.
For the pumps which work with the OBC controller, see the Operat-
ing instructions for the water valve connection.
For MT model