Typical connections
Optional electrical connections recommended
by the manufacturer (version M only)
El
ectrical connections (version MT only)
Minimum electrical connections
Legend:
Pump cable includes temperature sensor information
ATHM
or
ATPM
"ISOL." VALVE
Power supply
Start
Customer relay
K1
V3
V2
K1
3
Pump cable
THERMOST.
(pump
models MT)
2
Remote
control
Roughing
pump
High vacuum isolation valve
Controller
1
Wiring diagram (typical)
“WATER”
VALVE
2
PE
Purge
valve
N2
PURGE
VALVE
AIR
VALVE
Roughing
isolation valve
K1
Air
valve
V1
1
Connections depends on the controller model.
Refer to the controller operating instructions
2
Electrical connections (Only for MT models)
3
Recommended to connect a wire to the pump to the
potential earth even in perturbated electromagnetical
environment
Water
valve
In this installation, we use:
- a roughing isolation valve
V1
between the turbo pump and the
roughing pump;
- a high vacuum isolation valve
V2
between the turbo pump and
the chamber to be pumped;
- a relay
K1
, their contacts drive the valve
V1
and the roughing
pump power supply;
- the thermostatic option (MT pump model).
We recommend that you fit a separate earth (ground) conductor to earth
pump. Use an un-insulated braid or a separate insulated green/yellow
conductor with a minimum 9AWG (3mm2) conductor (to comply with US and
Canadian installation codes). Use a screw and a shakeproof washer fitted to
the earth hole from the pump to secure the earth conductor to the pump. The
impedance between the pump-body and the earth connection point must be
< 0.1 Ohm at 25A.
GB 03199 - Edition 03 - June 10
1/1
B 401
Typical electrical wiring diagram
adixen Vacuum Products - Operating instructions - ATH 2300 M/MT