Specifications
30
7)
±
50 V output
Linearity: 0 to
±
5 V, within
±
100 mV
±
5 V to
±
50 V, within
±
5%
Output impedance: 10 ohms or less
Output waveform: 7
µ
s or less both rising and falling
Excessive deviation: 10% or less
Maximum output current: 400 mA or over
Synchronized Output Signal
1) Output signal name
MAIN INTERVAL
SET CYCLES END
EXT TRIG
DELAY
INTERVAL
TRAIN END
2) Output signal
Amplitude: +5 V
Pulse width: approx. 500
µ
s
Rising: 3
µ
s or less
Output impedance: 1 kilohm or less
Language
English
Power Source
Voltage: AC 110 V, 220 V or 240 V
±
10 %
(Changed voltage range by the connector)
Frequency: 50/60 Hz
Power input: 70 VA
Environment
Main operating environment: Laboratory
Operating Environment
Temperature: 10 to 40
°
C
Humidity: 30 to 90%
Atmospheric pressure: 70 to 106 kPa
Storage Environment
Temperature: –20 to 60
°
C
Humidity: 10 to 95%
Atmospheric pressure: 70 to 106 kPa