Description of Parts
15
Name
K
Output selector key
(
±
50 V /
±
10 V)
L
to ISO
(Isolator output connector)
M
Output terminal
Function
When the key is pressed, the lamp lights and 5 times the normal voltage of 0 to
±
10 V is supplied to the output terminal.
This can be used for field stimulation.
(max. 0 to
±
50 V, 400 mA)
Additionally, key operation does not affect output to the isolator (to ISO
connector).
Drives the isolator when connected to the connector.
Pin No.
Signal
1, 2 +15V
3 –15V
4 SIGNAL OUT
5 Range data
6 E
7 Range data synchronizing clock
8, 9, 10 Void
[Fig.16]
As the output waveform is the same as that of the isolator output connector, this
is useful for monitoring the stimulation waveform.
It can also be used for the stimulation signal to the microelectrode amplifier or
membrane potential fixing amplifier .
However, when output is switched to
±
50 V, the waveform is not the same as
the isolator output connector signal, and five times the voltage is generated.
Use this terminal directly for field stimulation.
[Note] Take care to avoid short-circuiting during stimulation, otherwise
no stimulation will occur.