Acquisition modes
MSX-E3701-DIO
7.3
Common functions
The following functions are available both in Auto-refresh mode and in Sequence mode.
7.3.1
“Division factor”
Fig. 7-14: Acquisition: Division factor
In this section, the settling time is set, i. e. the time required to switch from one channel to another.
If only one channel is selected, this parameter is insignificant.
Select the division factor (value between 5 and 255).
7.3.2
“Acquisition time”
Fig. 7-15: Acquisition: Acquisition time
The total acquisition time is automatically computed according to the selected transducer and the
division factor. In Sequence Mode, the total time also depends on the selected delay mode (see
Chapter 7.2.3).
7.3.3
Trigger configuration
The acquisition can be started by an external signal.
The synchro trigger configuration has to be set both on the master’s and slave’s web interface.
Fig. 7-16: Acquisition: Trigger configuration