6243/44 DC Voltage Current Source/Monitor Operation Manual
6.4 GPIB in the TR6143 Mode
6-9
6.4
GPIB in the TR6143 Mode
Explained here are the GPIB commands, output data format, and status byte in the TR6143 mode.
6.4.1
TR6143 Mode GPIB Code List
In the TR6143 mode, the (1) TR6143 unique commands can be used but the (2) commands cannot be used.
Other commands can be used in both the Original Mode and the TR6143 mode.
(1) TR6143 mode unique commands
Table 6-4 shows the TR6143 commands that are not available in the Original Mode. The defaults in-
dicate the status of commands which can be initialized by factory default initialization, DCL, or C
command.
During sweep operation and DC/pulse mode operation, only commands with the symbol in Table
6-4 are accepted.
All the commands are accepted in the standby status.
Table 6-4 TR6143 Mode Unique Commands (1 of 2)
Item
Command
Description
Default
setting
Operation
6243/44
commands
Remarks
During DC/PLS
operation
During
sweep
operation
Source
Source function, source range,
and measurement function
settings
V3
Voltage source, 320 mV range, current measurement
V3
V4
Voltage source, 3.2 V range, current measurement
V4
V5
Voltage source, 32 V and 20 V ranges, current measurement
V5
V6
Voltage source, 110 V range, current measurement
V6 *4
*1
I-1
Current source, 32
A range, voltage measurement
I-1 *4
(TR6143 mode)
I0
Current source, 320
A range, voltage measurement
I0
I1
Current source, 3.2 mA range, voltage measurement
I1
I2
Current source, 32 mA range, voltage measurement
I2
I3
Current source, 320 mA range, voltage measurement
I3
I4
Current source, 2 A and 3.2 A ranges, voltage measurement
I4
I5
Current source, 10 A range, voltage measurement
I5 *5
Source mode
C1
DC mode
MD0
T0
DC sweep mode automatic trigger
MD2, ST0
T1
DC sweep mode external trigger
MD2, ST1
Source response
RP0
SLOW
-
*2
RP1
FAST
-
Measure-
ment
Auto zero
AC0
OFF
AZ0
AC1
ON
AZ1
Measure-
ment opera-
tion
Comparison operation result
buzzer
UZ3
The buzzer sounds when the comparison operation result is
HI.
BZ1
UZ4
The buzzer sounds when the comparison operation result is
GO.
BZ2
UZ5
The buzzer sounds with the comparison operation result is LO.
BZ3
Sweep
Sweep repetition number
T2
Repeats once
SS1
T3
Repeats continuously
SS0
Random sweep memory com-
plete setting.
C3
Used when setting the random sweep memory with the N
command.
(Example: N30, D1V, D2V, C3)
P
Содержание 6243
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 18: ......
Страница 22: ......
Страница 84: ...6243 44 DC Voltage Current Source Monitor Operation Manual 3 1 Menu Index 3 2 VM IM 3 3 3 19 VS IS 3 3 3 19...
Страница 106: ......
Страница 212: ......
Страница 238: ......
Страница 256: ......
Страница 280: ......