6243/44 DC Voltage Current Source/Monitor Operation Manual
9.2.1 Voltage/Current Source
9-14
Temperature coefficients:
At 0 to 50
C
Overall accuracy of reverse polarity limiters:
The accuracy of source value and reverse polarity limiters is obtained by
adding the data in the following table to the overall limiter.
(However, stability and temperature coefficients do not apply to the fol-
lowing table.)
Source linearity:
0.012% of range
However, the following errors are added for the 320 mA, 3.2 A, and 10 A
ranges.
Maximum output current:
10 A up to 7 V,
4 A up to 20 V
Maximum tracking voltage:
20 V up to 4 A,
7 V up to 10 A
Voltage source /
Voltage limiter
Range
Temperature coefficient
(ppm of s V)/
C
320 mV
3.2 V
20 V
15 + 30
V
15 + 50
V
15 + 500
V
Current source /
Current limiter
Range
Temperature coefficient
(ppm of s A + A
Vo/1 V)/
C
320
A
3.2 mA
32 mA
320 mA
3.2 A
10 A
25 + 10 nA + 100 pA
20 + 100 nA + 1 nA
20 + 1
A + 10 nA
20 + 10
A + 100 nA
20 + 100
A + 1
A
90 + 1 mA + 10
A
Vo:
Compliance voltage (0 V to
20 V)
Reverse polarity /
Voltage limiter
Range
Overall accuracy
(% of s V)
320 mV
3.2 V
20 V
0.25 + 8 mV
0.25 + 8 mV
0.25 + 80 mV
Reverse polarity /
Current limiter
Range
Overall accuracy
(% of s A)
320
A
3.2 mA
32 mA
320 mA
3.2 A
10 A
0.25 + 6.5
A
0.25 + 65
A
0.25 + 650
A
0.25 + 6.5 mA
0.25 + 65 mA
0.25 + 650 mA
Range
(% of setting)
Current source
320 mV
0.01
3.2 V
0.02
10 V
0.07