6243/44 DC Voltage Current Source/Monitor Operation Manual
2.2.4 Pulse Mode Measurement
2-20
Initializing the setting conditions
Initializing the 6243/44
1. Press the
MENU
key.
The parameter group selection screen appears.
2. Select
PARAMETER LOAD
in the parameter group by rotating the Data knob.
3. Press the key.
4. Select the
Load Default
parameter by rotating the Data knob.
5. Press the key.
6. Press the
DIRECT/ENTER
key.
Upon initialization completion, “Done” appears.
7. Press the
EXIT
key.
The measurement screen returns.
Connecting the DUT
Connect the DUT using the test leads and alligator clips.
8. Connect the alligator clips A08532 to the supplied test leads A01044.
9. Plug the test leads into the
HI OUTPUT
and
LO OUTPUT
terminals.
10. Clip the 1 K
resistor with the alligator clips.
Setting the Pulse value (source value)
11. Press the
DIRECT, 2, V,
and
ENTER
keys in order.
Setting the Current limiter
12. Press the
LIMIT
key.
The cursor moves to the limiter indication.
13. Press the
DIRECT, 3, mA,
and
ENTER
keys in order.
VS:+2.0000V
Cursor
L: 0500.0mA
Cursor
L: 3.0000mA
Cursor