6243/44 DC Voltage Current Source/Monitor Operation Manual
2.2.3 DC Measurement
2-18
20. Move the cursor with the key and set the voltage source value to
2 V
using the
Data knob.
The measured current value when 2 V is applied to the 1 K
resistor appears.
(Indicating the current value of operating point B shown in Figure 2-7 (c).)
21. Press the
POLARITY -
key to set the voltage source value to
-2 V
.
The measured current value when -2 V is applied to the 1 K
resistor appears.
(Indicating the current value of operating point C shown in Figure 2-7 (c).)
22. Press the
DIRECT, +/-, 4, V,
and
ENTER
keys in order.
The voltage source value becomes 4 V in the 32 V range. (4 V in the 20 V range
of the 6244.)
“
+lm
” appears because the current is restricted by the limiter.
(Indicating the current value of operating point D shown in Figure 2-7 (c).)
23. Press the
VM/IM
key.
The voltage between the HI-LO terminals appears.
(Indicating the voltage value of operating point D shown in Figure 2-7 (c).)
AZERO
IM+2.00000mA
VS:+2.0000V L: 3.0000mA *DC
AZERO
IM-2.00000mA
VS:-2.0000V L: 3.0000mA *DC
AZERO
IM+3.00000mA +Im
VS:+04.000V L: 3.0000mA *DC
AZERO
VM+0.30000V +Im
VS:+04.000V L: 3.0000mA *DC