6243/44 DC Voltage Current Source/Monitor Operation Manual
5.2.10 GPIB Code List
5-25
SWEEP
Linear sweep
SN
Specifying a unit: The optimal range is automatically set.
SN <data>UNIT, <data2>UNIT, <data3>UNIT
<data1>UNIT: The start value (The voltage or current value)
<data2>UNIT: The stop value (The voltage or current value)
<data3>UNIT: The step value (The voltage or current value) (The polarity is ignored.)
Units:
MV, V, UA and MA
• Only the sweep type is set when all setting values are omitted.
Each value cannot be omitted individually.
• An error occurs when an unused unit in the current source function is specified.
Default value: SN+000.01E-3V, +001.00E-3V, 000.01E-3V
SN?
The linear sweep query command
Responds with: SN
aaaaaaE
aUNIT,
bbbbbbE
bUNIT, ccccccE
cUNIT
aaaa: The start value
bbbb: The stop value
cccc: The step value (polarity is indicated by space.)
Unit: V or A
The voltage and current value responses depend on the current source function setting.
Log sweep
SG
Specifying a unit: The optimal range is automatically set.
SG <data>UNIT, <data2>UNIT, <data3>
<data1>UNIT: The start value (The voltage or current value)
<data2>UNIT: The stop value (The voltage or current value)
<data3>:
The number of the step split (1, 2, 5, 10, 25, and 50)
Unit:
MV, V, UA and MA
The number of the split: 1, 2, 5, 10, 25, and 50
• Only the sweep type is set when all setting values are omitted.
Each value cannot be omitted individually.
• An error occurs when an unused unit in the current source function is specified.
Default value: SG+000.01E-3V, +001.00E-3V, 10
SG?
The log sweep query command
Responds with: SG
aaaaaaE
aUNIT,
bbbbbbE
bUNIT, cc
aaaa: The start value
bbbb: The stop value
cc:
The number of the split (1, 2, 5, 10, 25, and 50)
Unit: V or A
The voltage and current value responses depend on the current source function setting.
Random sweep
SC
SC <data1>, <data2>
The SC start address and stop address
<data1>: The start address (0 to 4999)
<data2>: The stop address (0 to 4999)
• Only the sweep type is set when all setting values are omitted.
Each value cannot be omitted individually.
• The start address and stop address set during the operation can be changed while the
sweep operation pauses.
Default value: SC0000, 0000
SC?
The random sweep query command
Responds with: SCaaaa, bbbb
aaaa: The start address (0 to 4999)
bbbb: The stop address (0 to 4999)
The sweep type
query
SX?
Responds with
Linear sweep: The same response as the SN? command response.
Log sweep:
The same response as the SG? command response.
Random sweep: The same response as the SC? command response.
Bias value
SB
Unit Specification: The optimal range is set automatically.
SB <data>UNIT
Units: MV, V, UA and MA
An error occurs when an unused unit in the current source function is specified.
No unit specification: The current bias range is used.
SB <data>
Default value (6243): SB+000.00E-3V (Voltage function)
SB+00.000E-6A (Current function)
Default value (6244): SB+000.00E-3V (Voltage function)
SB+000.00E-6A (Current function)
SB?
The bias value query command
Responds with: SB
aUNIT
aaaa: Voltage or current source value
Units: V or A
The voltage and current value responses depend on the current source function.
Sweep range settings SR0
The sweep range automatic setting
SR1
The sweep range fixed setting
SR?
The sweep range setting query command.
Responds with SR...
Table 5-5 GPIB Command List (5 of 9)
Item
Command
Description
Default
Operation
During DC/
PLS operation
During sweep
operation