© Adam Equipment Company 2011
5
2.
SPECIFICATIONS
INPUT SECTION
Load Cells
Up to 4 , 350 ohm load cells
Minimum 87 ohms, maximum 1120 ohms
Connection
6 wires
2 excitation, 2 sense, 2 signal
Excitation
5Vdc
Sensitivity
0.15uv/d (GK-M, 1.5uv/e)
Linearity
0.01% FS
Zero Range
0- 10mv
Signal range
0-40mv
ADC Sensitivity
Approximately 0.02 µv/ADC count
DIGITAL SECTION
Maximum Range
Typically 1kg – 30000kg
Divisions
Up to 30,000, (GK –M, 3000 or less)
Weigh units
kg / g for Europe, Asia, Africa
kg / g / lb / oz / lb:oz for USA
kg only for GK-M series
Stabilisation Time 2 Seconds typical
Operating
Temperature
-10°C - 40°C
14°F - 104°F
Power supply
230 VAC 50/60 Hz
12V 800ma adaptor for USA verisons
Battery
Internal rechargeable battery
Calibration
Automatic External
Display
6 digits LCD digital displays
with capacity tracker and symbols for units
Indicator Housing ABS Plastic
Overall
Dimensions
(wxdxh)
260 x 170 x 115mm
10.2” x 6.7” x 4.5”
Net Weight
2 kg / 1 lb
Applications
Weighing and check weighing
Functions
Weighing, Check Weighing, Parts counting, check-counting, , Animal
Weighing, Accumulating memory,
Interface
RS-232 bi-directional interface
English, German, French, Spanish selectable text
For approved indicators the input specifications is limited to 1.5 µv per division and
the number of divisions is limited to 3000d. Kilograms only.