55 © Adam Equipment Company 2011
Press
[Tare]
to accept the value.
Press
[Zero]
to return to weighing.
16.2.6
F6 -SUCCESSIVE TARE
To enter this parameter, press the
[Tare]
key when “
F6 SCS”
is shown.
The display will show if the successive tare is on or off.
Press the
[Func]
key to change the value.
Press
[Tare]
to accept the displayed value.
Press
[Zero]
to return to weighing.
16.2.7
F7 –ADC COUNTS
To enter this parameter, press the
[Tare]
key when
“F7 Cnt”
is shown.
This parameter allows you to view the A/D counts from the internal A/D converter.
This can be an aid to service.
Press the
[Tare]
key to return to the PARAMETER menu.
Press the
[Zero]
key to return to weighing.
Typical value at zero is 30,000-90,000 (approx.)
Typical value at full capacity is 500,000 (approx.)