3. 

System installation

1. Extension coding (take one extension as an example)

Connect a pair of white and black wires of the extension or waterproof switch onto the two-core bus corresponding 
to the mainframe 1 and COM terminals according to the illustration (please pay attention to the polarity).

Connect a pair of white and black wires of the extension or waterproof switch onto the two-core bus corresponding 
to the mainframe 1 and COM terminals on the mainframe according to the illustration, connect a pair of yellow and 
black wires on the rear of the extension to the two-core wire in serial, and then connect to the negative and positive 
poles of the external power supply (please pay attention to the polarity), plug the transfusion alarm device onto the 
transfusion alarm port. (When the number of extensions is within 30, only one additional external power supply is 
required, when the number of the extensions goes beyond 30, two additional power supplies are required).

A. If broadcast and transfusion alarm functions are not required for the extension, only a two-core bus is required 
for connection between the extension and the mainframe. For specific wiring method, please refer to the following.

B. If broadcast and transfusion alarm functions are required for the extension, a four-core bus is required for 
connection between the extension and the mainframe. For specific wiring method, please refer to the following.

Preparation for extension online coding: Connect one pair of white and black wires on the extension or the 
waterproof switch respectively onto the two-core bus corresponding to 1 and COM terminals of the mainframe. 
For the specific wiring, please refer to the 'Wiring mode for extensions, waterproof switches and mainframes' 
of article 2 below (for details, please refer to P6).

Extension online coding setup: Connect the extension with the mainframe according to the above-mentioned 
method, press down the mainframe's power switch for about two seconds, and the mainframe starts up and enters 
the self check status. After self check, press the function key on the mainframe for about two seconds, and enter 
the system setup options. For specific procedures of online coding setup, please refer to the 'Online coding setup' 
of article 11 in the 'System setup' in the fifth time (for details, please refer to P15).

2. Connection between extensions, waterproof switches and mainframes

Mainframe rear diagram

2P

/

2

Waterproof 

switch 

2P

/

Connected to a pair of yellow and black 
wires on the extension (note the polarity)

2

P /

External power 
supply(14V 2.5A)

Negative pole

Positive pole

2

P /

6

Mainframe rear diagram

External power 
supply(14V 2.5A)

Negative pole

Positive pole

Connected to a pair of yellow and black 
wires on the extension (note the polarity)

Waterproof 

switch 4

Waterproof 

switch 60

Extension 1

Extension 3

Extension 59

2

Waterproof 

switch 

Waterproof 

switch 4

Waterproof 

switch 60

Extension 1

Extension 3

Extension 59

Содержание HC-30

Страница 1: ...ystem profile Technical specifications System installation System setup Operation instructions Matters for attention Troubleshooting Service commitment Hospital nursing intercom system installation an...

Страница 2: ...e for example alarmsound flashing light and voice announcement XX transfusion completed Three level nursing Threelevels of nursing namely special high and common can be seton the mainframe according t...

Страница 3: ...ht 2 The mainframe haveChinese LCD displaywith menu functionsand operations asa result theoperation is more convenient and simpler and theproduct is ableto display variousstatuses date andtime Only on...

Страница 4: ...to callthe mainframe There is atransfusion alarm interfaceon the extensionthat can cooperatewith the transfusionalarm device When transfusion is completed the extension reportsalarms to themainframe a...

Страница 5: ...talled inthe corridor tofacilitate nursing personnelin the corridorin observing The dot matrixdisplay can displaythe calling extensionnumber and nursinggrade in caseof no call it is ableto display yea...

Страница 6: ...th thewireless receiver When an extensioncalls the wireless transmitter will automaticallytransmit the callingextension number tonursing personnel equipped withthe wireless receiver Generally installe...

Страница 7: ...pecificwiring method pleaserefer to thefollowing Preparation for extensiononline coding Connectone pair ofwhite and blackwires on theextension or the waterproof switch respectivelyonto the two corebus...

Страница 8: ...e 5 Connection betweenthe door lampand extension take theextensions inside thesame ward as an example 1 Connect thered and blackwires of thedoor lamp ontothe red andblack wires ofthe extension accordi...

Страница 9: ...n the mainframe Corridor display 3P 2P 8 Connection betweenmonitoring statistics lineconcentration box monitoringstatistics software computer andmainframe A Insert the power cord into the line concent...

Страница 10: ...tomatically detectextensions The mainframechecks extensions oneby one fromextension 1 theextension is normal andthe indicator lightblinks once ifthere is anyextension wrong orthe channel isnot connect...

Страница 11: ...nursing station Mainframe Level 2management nursing station Power source Line concentration box Corridor display Power source Power source Power source To list of patients To extension outside power s...

Страница 12: ...vel is 01 four panes ofvolume for theday twopanes for volumeat night andfour panes for hands free talkvolume 4 The startupalarm time is00 00 note thisfunction setup isnot activated andthe mainframe fo...

Страница 13: ...and there isa voice announcement reporting the extensionnumber and calllevel To remove thehigh level nursingsetup for anextension just select the routing switchof the channelin the specialnursing set...

Страница 14: ...Call VOL FUNCTION MNTR PWR FUNCTION MNTR ALL FUNCTION MNTR PWR FUNCTION MNTR ALL Volume FUNCTION MNTR Press the or key to enter setup to enter setup Press or key to select the volume Press or key to s...

Страница 15: ...odel HC After the previous step press the Function key to enter the mainframe level setup Tiering Setup Model HC 10 Ward bed setup takeextension 13 setto bed 03of ward 204as an example Sickbed Setup M...

Страница 16: ...vious item press the Function key to enter the forwarding upward duration setup Tier Transfer Setup FUNCTION MNTR ALL Model HC PWR 11 Online coding setup e g code theextension as No 1 extension After...

Страница 17: ...at the sametime for groupcall Every five channelsform one group namely 1 5 6 10 11 15 as one group Tocall a groupin group call press down the firstchannel of routingswitch of thegroup for example 1 6...

Страница 18: ...ompanywill try itsbest to meetyour needs This product canonly be maintainedby our company strained professional personnel The rear coverof the mainframe shall notbe opened withoutour company s authori...

Страница 19: ...e key is pressed Unable to operatenormally after startup The extension cancall the mainframe however it is difficult for themainframe to choose anextension Bus disconnection 1 Extension damaged 2 Conn...

Отзывы: