4-68
4.2 Operation and features
SPM-32A ... SPM-36A
2nd setup:
3rd setup:
The whole memory can be examined quickly in this way.
Display mode is switched off by entering:
The instrument is set to the last setup and starts measuring.
4.2.8
Avoiding errors due to large common-
mode voltages
Use screened cables if stray fields are present, otherwise
induced interference voltages may result!
Stray fields are found in the areas surrounding high tension
lines for example. Spurious signals will be induced in a twin
core cable connected to the balanced input of the instrument
in such cases, resulting in an incorrect measurement due to
the finite common-mode rejection ratio of the input circuit.
These errors can be minimised by either
– Use a CF cable with good shielding and connect it to the
test ground connection of the system under test.