12. 3 FUSE REPLACEMENT
P
IEZOTOME
2 is protected by two fuses located in the power receptacle (Fig. 2-1).
The replacement procedure is as follows:
- Switch the device OFF (position O).
- Unplug the power cord from the electricity supply.
- Unplug the power cord from the receptacle (Fig. 2-1). Insert the tip of a flat screwdriver into the
notch above the fuse drawer to release it.
- Remove the old fuses.
- Replace the old fuses with fuses of the same type and rating.
- Push the fuse drawer back into its housing until it clicks to indicate the correct position.
- Plug the power cord into the receptacle (Fig. 2-1). Plug the power cord into the electrical supply.
Note
:
The device also has an internal fuse not accessible to the user.
Contact SATELEC after-sales service (see Chapter 2 - Repairs).
All the information necessary for the repair of faulty components is made available by SATELEC to the
technical personnel of the SATELEC-approved dealer network who carry out the work.
12. 4 OPERATION FAULTS
Refer to the table below:
Fault
Possible cause
Remedy
Device does not operate (LCD
screen off)
Power cord not plugged in properly.
Check the electric socket.
Return to SATELEC after-
sales.
Power switch in position O (OFF).
Place the power switch in
position I (ON).
No power voltage.
Call an electrician.
Power receptacle fuse(s) blown.
Replace the power receptacle
fuses.
Internal fuse blown
Return to SATELEC after-
sales.
Device does not operate (LCD
screen on)
Transmission fault.
Switch the device OFF then
ON again.
Return to SATELEC after-
sales.
No spray
Irrigation solution bag or bottle
empty.
Replace the bag or bottle of
irrigation solution.
Irrigation off.
Press the irrigation ON/OFF
button.
Irrigation line clogged.
Change irrigation line.
Inadequate spray
Incorrect irrigation flow rate
adjustment.
Adjust the irrigation flow
rate.
Power fault
weak tip vibration
Tip worn or distorted.
Replace the tip.
Incorrect power adjustment.
Refer to the TipBook for the
necessary information.
Incorrect use: incorrect angle of
attack or inadequate pressure on the
tooth.
Refer to the TipBook for the
necessary information.
Liquid or moisture present between
handpiece and cord.
Dry the electrical contacts
thoroughly.