f) Technical characteristics
Manufacturer
: SATELEC
Device name
: P
IEZOTOME
2
Electrical power supply
:
- Voltage: 100 VAC to 230 VAC.
- Frequency: 50 Hz/60 Hz.
- Rated power: 150 VA at 230 VAC.
P
IEZOTOME
function
Operation
:
Intermittent service: 10 min. ON/5 min OFF
Output characteristics
:
No-load voltage: 250 volts (nominal value without handpiece)
Min. ultrasonic frequency: 28 kHz
Irrigation flow rate: 10 to 120 ml/min. (nominal value)
Adjustment in 10 ml/min. steps (nominal value)
Flush flow rate: 120 ml/min. (nominal value)
N
EWTRON
function
Operation
:
Intermittent service: 10 min. ON/5 min. OFF
Output characteristics
:
No-load voltage: 150 volts (nominal value without handpiece)
Min. ultrasonic frequency: 28 kHz
Irrigation flow rate: 10 to 40 ml/min. (nominal value)
Adjustment in 1 ml/min. steps (nominal value)
Flush flow rate: 120 ml/min. (nominal value)
Protection
Electrical equipment class: Class 1
Electrical safety class:
Type BF in P
IEZOTOME
mode
Type BF in N
EWTRON
mode
Safety devices
Safety shut off if internal malfunction
2 fuses (power receptacle): 5 mm x 20 mm/2 AT for 100 VAC to 230 VAC
1 internal fuse not accessible to the user reference F1: 5 mm x 20 mm – 10 AT/250 VAC
Overall dimensions
Control unit
:
Width: 472.9 mm
Height: 149.5 mm
Height with bracket: 471.1 mm
Depth: 339.9 mm
Weight: 5 kg without accessories
Footswitch
:
Width: 311 mm
Depth: 209 mm
Height: 181 mm