P
IEZOTOME
function
- Connect the P
IEZOTOME
2 handpiece cord to one of the connectors of the device (Fig. 1-7 and 8).
- Screw the selected tip onto the handpiece using a tip wrench (according to the recommendations of
the clinical booklets).
- Fasten the irrigation line to the cord using line clips.
- Connect the end of the irrigation line to the P
IEZOTOME
2 handpiece.
- Prime the irrigation circuit by pressing the flush/prime button on the screen or the multifunction
footswitch.
- When the irrigation solution reaches the end of the tip, release the footswitch.
- Use the P
IEZOTOME
2 in accordance with good dental practice.
N
EWTRON
function
- Check that there are no traces of moisture on the N
EWTRON
handpiece or the scaler cord connections.
Remove any moisture by wiping and by drying with the multifunction syringe fed with medical-quality
filtered air.
- Connect the scaler cord to one of the connectors of the device (Fig. 1-7 and 8).
- Align the electrical contacts of the N
EWTRON
handpiece and the scaler cord, then plug the handpiece
into the cord.
- Screw the selected tip onto the handpiece using a tip wrench (according to the
T
IP
B
OOK
recommendations).
- Fasten the irrigation line to the cord using line clips.
- Connect the end of the irrigation line to the scaler cord.
- Prime the irrigation circuit by pressing the flush/prime button on the screen or the multifunction
footswitch.
- When the irrigation solution reaches the end of the tip, release the footswitch.
- Use the P
IEZOTOME
2 in accordance with good dental practice.
X - -SHUTTING DOWN THE DEVICE
Upon completion of the dental procedure:
- Remove the bottles or bags of physiological saline or sterile water from the brackets.
- -Remove the irrigation line perforating pins from the bottles or bags.
- Immerse the irrigation line perforating pins in a vessel containing distilled water.
- Rinse each irrigation line alternately, along with the handpiece, by running the flush function until the
vessel and the irrigation lines are completely empty.
- Remove the irrigation line clips.
- Disconnect the single-use irrigation lines from the handpieces and dispose of them in a safety
container for soiled medical items.
- Disconnect the P
IEZOTOME
2 handpiece or the N
EWTRON
handpiece and the scaler cord.
- Switch the device OFF (O).
XI - -ROUTINE MAINTENANCE / STERILIZATION
Important:
The device must be switched OFF during cleaning and disinfection procedures.
The only irrigation lines that can be sterilized are those identified as sterilizable.
During sterilization, items made of different metals must not come into contact. Any contact would
establish galvanic couples and cause localized damage.