USER MANUAL 16-2045
Rev. 09/2017
Page 13 of 25
6.7 Parameters
The instantaneous flow is shown on the screen in litres/minute.
The volume of gas used is shown in litres on the screen (accurate to the decilitre. For
illustrative purposes only.)
This measurement begins at the same time as insufflation and is only reset to zero when it
reinitializes each time the device is started up (or from the parameters interface menu).
If necessary, set it to zero using the menu.
Only use medical-grade CO2, in accordance with European standards.
6.8 End of operation
Halt insufflation by tapping
the “STOP” button. The chronometer stops.
Immediately disconnect the tubing hoses between the trocar and the insufflator to avoid any
liquid or gas backflow into the device.
The tubing hoses used must be discarded after use in an appropriate container.
The total volume of gas insufflated during the operation is displayed; reset the gas counter to
zero if another operation is planned without turning off the device.
Press the Standby button [S1] to put the device in Standby mode. Turn the power switch [S2]
to the “O” position to turn off the device.
7 Special features
7.1 Control from the Ackermann Fusion camera head (Ackermann Fusion
camera only)
The following insufflator functions can be controlled using one of the three programmable
buttons located on the camera head: RUN (in low flow mode), HIGH FLOW RATE, STOP. A
special communication cable must connect the Ackermann Fusion camera to this insufflator
at the dedicated socket [C5] (see Ackermann Fusion camera user manual for further
information on configuring this switch). In addition, this communication cable is for receiving
feedback information (instantaneous pressure in the cavity, warning messages) on the
surgical monitor.
7.2 Autonomous operation with the CO² cylinder
This patented function indicates the insufflation time remaining based on the remaining CO2
capacity in the cylinder. It helps the surgical team to better manage:
- surgical operation time,
- cylinder changes.
The autonomous function is triggered when the pressure of the CO2 cylinder is below 33 bars
and when autonomy is equal to or less than 99 minutes.