Note:
If the insufflator does not switch on, this could be due to damage to the fuses. If so, it is
advisable to turn off the power, check and if necessary replace the fuses (use only T25AL
–
250 V delayed-action UR fuses)
For any other problems, contact your nearest after-sales service department.
11 Technical characteristics
Pneumatics:
•
CO2 gas supply system: US 7/16
˝
connector
•
Pressure range:
- CO2 cylinder: 10 to 60 bars
- central gas wall point outlet: 3 to 5 bars.
•
Maximum flow rate
- 45 l/min in high-flow mode
- 2 l/min in low-flow mode
•
Pressure setting: 0 to 25 mm Hg (accuracy: 1 mm Hg)
•
External exsufflation valve
•
Automatic low-flow / high-flow function selection option
Interface:
•
Touch screen displaying: immediate flow rate, pressure in the cavity, total volume
of CO2
used and low cylinder levels
Energy supply:
•
Types of energy: 100 - 230 V AC 50 - 60 Hz
•
Fused protection: 2 x 2.5 AT - 250 V delayed-action UR fuses
•
Power consumption: 75 VA
Mechanics:
•
Dimensions (W x H x D): 310 x 136 x 385 mm
•
Weight: 8000g
Operation - transport and storage environment:
•
Operating temperatures b 10
º
C and + 40
º
C
•
Operating relative humidity between 30 and 75 %
•
Transport and storage temperatures between - 10
º
C and + 45
º
C
•
Transport and storage relative humidity between 20 and 85 %
•
Operating, transport and storage atmospheric pressure between 700 hPa and
1060 hPa