USER MANUAL 16-2045
Rev. 09/2017
Page 10 of 25
5.4 Disconnecting the CO² cylinder
Before unscrewing the high-pressure hose:
- Check that the valve on the cylinder is closed, then:
• Lower the pressure inside the system by slowly unscrewing the tubing hose.
OR
• Use the Purge function (on the interface, tap PURGE) after closing the valve on the bottle.
5.5 Disconnecting from the wall point
Manually disconnect the low-pressure hose from the wall point.
5.6 Connecting to the Ackermann Fusion Camera
Turn off the devices.
Connect the cable supplied with the camera to the socket on the back side of the camera
control unit, and then to the back side of the insufflator [C5].
Turn on the devices.
6 Operating guidelines
This insufflator in only intended for use in diagnostic procedures or laparoscopic operations.
Any use outside of these areas constitutes improper use of the product and the user will
therefore be considered responsible for such use. The manufacturer does not accept any
liability in this case. This device must only be used by qualified staff. The surgeon and
anaesthesiologist shall at all times be responsible for the device and the anaesthesiologist
shall undertake special vigilance of gas levels in the blood. The safety features on this device
are in no way intended to reduce the responsibility of medical staff with respect to constant
attention to the screen and continuous monitoring of the patient.
6.1 Activation
Turn the switch [S2] to position
“
1
”
; the device is now in standby mode (the indicator light
on the Standby button [S1] will be flashing, and the Ackermann logo should appear on the
touch screen [L1]);
Press the Standby button [S1] or tap the touch screen to start the device (the indicator light
comes on and the calibration phase, which lasts for a few seconds, starts). Pressing the
Standby button again puts the device back in standby mode.
Note: If the insufflator is connected to the Ackermann Fusion camera, activating the camera
will automatically activate the insufflator and vice-versa. However, the insufflator can be put
in standby mode, or turned on, independently.