USER MANUAL 16-2045
Rev. 09/2017
Page 18 of 25
10 Troubleshooting
Warning
Possible Cause
Measures to take
Orange cylinder logo
Cylinder pressure at 40 bars
Change the cylinder. it is
advisable to prepare a
replacement cylinder as soon
as the cylinder pressure
reaches 40 bars.
gas cylinder valve closed
Open the valve
high-pressure hose connector
on the insufflator disconnected
Turn off the device, check that
the CO2 gas is flowing and turn
the device back on
Excessive pressure
sporadic action on the
pneumoperitoneum
none.
- if the pressure exceeds the
pressure set by 2 mm hg, the
device will create mild
exsufflation.
Tubing hose tangled
stretch out the tubing hose
Trocar valve closed during
insufflation
Open the trocar
Chck the tubing hose
Tubing hose missing
Connect the tubing hose
The tubing hose is not correctly
connected
Correctly connect the tubing
hose to [C1] and [C2]
Tubing hose fault detected
Sent the device to the
distributor for repair
Tubing hose obstructed
Bent tubing hose
stretch out the tubing hose
Trocar valve closed during
insufflation
Open trocar valve
CO2 not detected
gas cylinder valve closed
Open the valve
high-pressure cylinder or
insufflator hose disconnected
Correctly connect the hose to
the insufflator and to the
cylinder
CO2 gas wall point system
faulty
Check whether the pressure
supplied by the CO2 gas wall
point system is higher than 3
bars.
insufflator hose connection /
wall point system connection
Carefully connect the hose to
the CO2 gas wall point outlet
and to the insufflator