8 Suggested decontamination procedures
The insufflator is a medical device, not heat resistant and it cannot withstand immersion;
consequently it should be disinfected with a non-woven medium saturated disinfectant
detergent.
Decontamination methods and/or selected tools, it remains under the full responsibility
of the staff concerned.
This device cannot be autoclaved.
Existing alkaline solutions for the disinfection of certain medical devices are NOT
RECOMMENDED for the disinfection of this device.
Always disconnect the device before cleaning it.
After each use:
Discard disposable sterilized tubing hoses; do not attempt to re-sterilse them.
Clean up all possible spatters on the insufflator by removing them with a slightly
moistened cloth.
The device must always be decontaminated before sending it back to the manufacturer
or distributor.
9 After-sales service and maintenance
No particular maintenance operation is needed for this device.
Note: misuse is not covered by the warranty. If a fault persists and the device has to be
returned to the distributor, ensure that it is shipped in its original packaging.
Similarly, it is advisable to return the device in its entirely (control unit and power cables).
Kindly attach to the shipping order a short explanatory note about the fault detected.
The equipment must be disinfected before returning for repair. When returning the
equipment, check its condition and make notes on the delivery note if necessary, confirming
them with the carrier by registered letter as soon as possible.
Please contact your distributor who will guide you through the return process.