ProScale M150, M250 with LCD Digital Displays
Page 31 of 48
Pr14 – Output Signal Mode
[
0
, 1, 2]
Configures the hardware output signal for activation on MONitor drift
conditions or Upper/Lower limit alarm conditions.
Applies to 24VDC Displays only.
SEE Pr15, Pr16, Pr17 ALSO
0 = Off
1 = Monitor Drift
2 = Upper/Lower Limit
Pr15 – Output Polarity
[
0
, 1]
Used to configure the signal output.
N/O or N/C in relation to circuit ground.
0 = N/O, the output is Normally Open (not conducting to ground).
1 = N/C, the output is Normally Closed (conducting to ground).
Active only when Pr14=1
Pr16 – Lower Limit
[-999.999 to 999.999in] or [-9999.99 to 9999.99mm]
Sets the lower limit alarm value.
Default = 0.000
in.
(0.00mm)
Active only when Pr14=2
.
Default is set in Inches
.
Pr17 – Upper Limit
[-999.999 to 999.999in] or [-9999.99 to 9999.99mm]
Sets the upper limit alarm value.
Default = 5.000
in.
(127.00mm)
Active only when Pr14=2
.
Default is set in Inches.
Pr18 – Drift Tolerance
[.001 to 999.999in] or [.01 to 9999.99mm ]
Range of motion allowed (+ or -).
Default =. 010 in.
(.254mm)
while in MONitor mode
Default is set in Inches
Pr19 – Automatic Monitor ON Time
[
0
, 1 or 2]
Configures Display to automatically activate MONitor mode after
30 or 60 seconds of ReadHead inactivity.
0 = disabled
1 = 30 seconds
2 = 60 seconds
Pr20 – Automatic Monitor OFF Enable
[
0
, 1]
Configures Display to automatically exit MONitor mode after a
programmed distance (Pr21) has been exceeded from the drift
tolerance position (Pr18).
0 = disabled
1 = enabled
Pr21
–
Automatic Monitor OFF Distance
[0.001 to 999.999in] or [0.01 to 9999.99mm].
The distance that must be exceeded from the drift tolerance
position (Pr18) to deactivate monitor mode.
Active only when Pr20=1
Default = 0.500 in.
(12.70mm)
Default set in Inches