Calibration is recommended after initial installation and each time the balance/scale is moved.

Features

Calibration/adjustment can be performed only when:

– there is no load on the balance/scale,
– the balance/scale is zeroed, and
– the internal signal is stable.

If these conditions are not met, an error message 
is displayed.

Example:

Calibrate/adjust span of the balance/scale (here: model VIC-5101)

Step

Key (or instruction)

Display

1. Switch on the balance/scale 

(ON/OFF)

2. Zero the balance/scale

(ZERO)

3. Start calibration

(CAL)

The preset calibration weight is displayed without 
the weight unit (in this example, 5000 g)

4. To select a different calibration weight value

(F) repeatedly

5. Confirm calibration weight value and 

(ENTER)

start calibration/span adjustment

After the zero point is stored, the required
calibration weight flashes on the display.

6. Place the required calibration 

weight on the balance/scale

The readout stops flashing if the weight is 
applied within the defined time limit and tolerance.
If the weight value is accepted, the display stops 
flashing and the stability symbol 

appears 

on the display.

7. Remove the calibration weight

8. Calibration has been completed

Calibration/Span Adjustment

The weight required for calibration/adjustment is displayed.
Standard calibration weights for selected models: to remove,
see instructions on page 4.
Press (F) to select a different weight value.
To cancel the procedure: press and hold the (ENTER) key 
(> 2 sec.).

Содержание ACCULAB VICON VIC-123

Страница 1: ...de d emploi Istruzioni per l uso Manual de instrucciones ACCULAB VICON Electronic Precision Scales Balances Elektronische Pr zisionswaagen Balances lectroniques d analyse et de pr cision Bilance elett...

Страница 2: ...ies Options 16 C Marking 2 Contents Warnings and Safety Precautions Safety Information To prevent damage to the equip ment please read these operating instructions carefully before using your balance...

Страница 3: ...adjustable feet Storage Do not stack more than 3 balances on top of one another at a time Installation Choose a location that is not subject to the following nega tive influences Heat heater or direct...

Страница 4: ...lance scale on its side Open the battery compartment remove the compartment cover Install the battery in the compart ment Make sure the polarity is correct Close the battery compartment slide the cove...

Страница 5: ...off automatically unless a key is pressed Description of the Keys ON OFF On off key switches the balance scale on and off or switches it to the standby mode Battery operation on turns backlight on of...

Страница 6: ...e on balance scale 7 Toggle weight unit F Menu code Unit Conversion factor Display 1 user Grams 1 00000000000 o 2 grams factory setting Grams 1 00000000000 g 4 carat Carats 5 00000000000 o 5 Pound Pou...

Страница 7: ...ale 5 Zero the balance scale ZERO 6 Place reference sample quantity 20 on the balance scale 7 Select reference sample quantity in increments of 1 1 2 3 99 or F repeatedly briefly or in increments of 1...

Страница 8: ...mpty container on the balance scale 5 Tare the balance ZERO 6 Place the reference weight for 100 on the balance scale 7 If desired change the number of decimal places displayed 100 0 100 00 or 100 fac...

Страница 9: ...n the display application is active 4 If necessary zero the balance scale ZERO 5 Place oversized sample on balance scale 6 Start application program F Symbol flashes on the display the weight value is...

Страница 10: ...yed ENTER steadily symbol flashes 7 Remove sample from balance scale 8 Place the next sample on the balance scale in this example 575 g 9 Store value in memory Totalized stored weight ENTER is display...

Страница 11: ...ance scale ZERO 5 Start application program F 6 Confirm the display airval ENTER 7 Determine weight of the sample in air place sample on the balance scale 8 Store weight in air value ENTER 9 Remove sa...

Страница 12: ...different calibration weight value F repeatedly 5 Confirm calibration weight value and ENTER start calibration span adjustment After the zero point is stored the required calibration weight flashes on...

Страница 13: ...ons 1 2 Application filter 1 2 1 Final readout 1 2 2 Filling 1 3 Stability range 1 3 1 1 4 digit 1 3 2 1 2 digit 1 3 3 1 digit 1 3 4 2 digits 1 3 5 4 digits 1 5 Calib adjust linearization 1 5 1 Calibr...

Страница 14: ...play calErr Calibration parameter not met e g Calibrate only when zero is displayed Balance scale not zeroed Press ZERO to tare the balance scale Balance scale is loaded Unload the balance scale appEr...

Страница 15: ...of operation w 9V battery Alkaline approx h 11 11 14 Rechargeable fully ch NiMH avg h 2 5 2 5 4 Model VIC 5101 VIC 3101 VIC 1501 VIC 711 VIC 511 VIC 10KG VIC 6KG VIC 4KG Weighing capacity g 5100 3100...

Страница 16: ...request Accessories Options C Marking The balance scale complies with the following EC Directives and European Standards Council Directive 89 336 EEC Electromagnetic compatibility EMC Applicable Europ...

Отзывы: