To configure the balance/scale; i.e., adapt the balance/scale to individual requirements.
Step
Key (or instruction)
Display
1. Switch off the balance/scale
(ON/OFF)
2. Switch on the balance/scale and
(ON/OFF)
while all segments are displayed
(ZERO) briefly
Navigation in the Setup Menu
Key
Press briefly
Press and hold
(ENTER)
Menu level: move to the right (cyclical)
Confirm setting
(F)
Menu item: Scrolling
–
(ZERO)
Menu level: Move to the left
Save settings and exit Setup
Parameter Settings (Overview)
ο
Factory setting
√
User-defined setting
Setup
1
Weighing
1.1. Adapt filter
1.1.1
Very stable conditions
menu
1.1.2
ο
Stable conditions
1.1.3
Unstable conditions
1.1.4
Very unstable conditions
1.2. Application filter
1.2.1
ο
Final readout
1.2.2
Filling
1.3. Stability range
1.3.1
1/4 digit
1.3.2
1/2 digit
1.3.3
1 digit
1.3.4
ο
2 digits
1.3.5
4 digits
1.5. Calib./adjust./linearization:
1.5.1
ο
Calibration/adjustment
Function of the (CAL) key
1.5.2
Linearization: for service personnel only
1.5.3
Key blocked
1.6. Auto zero
1.6.1
ο
On
1.6.2
Off
1.7. 1st weight unit, or 2nd unit in
1.7.1
User-defined unit; see
Toggle Weight Units app.
to 1.7.23
“Toggling between Weight Units”
5. and 6.
Only relevant with built-in data interface: see corresponding interface description
8. Additional
8.1. Block key functions
8.1.1
All keys blocked except for (ON/OFF)
functions
and (ZERO)
8.1.2
ο
All keys unblocked
8.2. Automatic shut-off
8.2.1
ο
After 2 minutes
8.2.2
After 5 minutes
8.2.3
After 10 minutes
9. Reset menu
9.1. Factory settings
9.1.1
Restore
9.1.2
ο
Do not restore
Configuration (Setup Menu)