background image

5

Basic Weighing Function

Features

– Zeroing the balance

You can zero the balance/scale within the entire weighing
range, up to the maximum capacity.

Preparation

§

Switch on the balance/scale: press the (ON/OFF) key

§

If necessary, zero the balance/scale: press the (ZERO) key

$

If necessary, change the configuration settings: 
see the chapter entitled “Configuration”

$

If desired, load the factory settings: 
see the chapter entitled “Configuration,” menu item 9.– 1

Additional functions:

$

Switching off the balance/scale: press the (ON/OFF) key

$

Battery operation: automatic shut-off after 2,5 or 10 min-
utes; see chapter on configuration. Example: 2 minutes.
If the weight readout remains unchanged and no keys are
pressed for at least two minutes, the battery symbol

“ starts flashing. After another 2 seconds, the 

balance shuts off automatically, unless a key is pressed.

Description of the Keys

(ON/OFF)

On/off key: switches the balance/scale on and
off or switches it to the standby mode.
Battery operation: on; turns backlight on; off

(ZERO)

Zeros the balance; press and hold 2 seconds:
opens the application menu

(CAL)

Starts calibration/adjustment

(F)

Starts an application program; 
Scrolling in application menu, configration
menu and calibration menu

(ENTER)

Confirms the selected setting;
Exits application, configuration & calibration
menu if key is pressed and held for more than 
2 seconds.

(PRINT)

Generates a printout or data output

Operation

Example:

Determine weight of sample

Step

Key (or instruction)

Display

1. Switch on the balance/scale

(ON/OFF)

Self-test runs
Display: Software version

r 31.01 

2. Open the in-use cover and leave open

while weighing

3. Place container on the balance/scale

52.0

g

(in this example, 52 g)

4

. Zero the balance/scale

(ZERO)

0.0

g

5. Place sample in container on balance/scale

150.2

g

(in this example, 150,2 g).

Содержание ACCULAB VICON VIC-123

Страница 1: ...de d emploi Istruzioni per l uso Manual de instrucciones ACCULAB VICON Electronic Precision Scales Balances Elektronische Pr zisionswaagen Balances lectroniques d analyse et de pr cision Bilance elett...

Страница 2: ...ies Options 16 C Marking 2 Contents Warnings and Safety Precautions Safety Information To prevent damage to the equip ment please read these operating instructions carefully before using your balance...

Страница 3: ...adjustable feet Storage Do not stack more than 3 balances on top of one another at a time Installation Choose a location that is not subject to the following nega tive influences Heat heater or direct...

Страница 4: ...lance scale on its side Open the battery compartment remove the compartment cover Install the battery in the compart ment Make sure the polarity is correct Close the battery compartment slide the cove...

Страница 5: ...off automatically unless a key is pressed Description of the Keys ON OFF On off key switches the balance scale on and off or switches it to the standby mode Battery operation on turns backlight on of...

Страница 6: ...e on balance scale 7 Toggle weight unit F Menu code Unit Conversion factor Display 1 user Grams 1 00000000000 o 2 grams factory setting Grams 1 00000000000 g 4 carat Carats 5 00000000000 o 5 Pound Pou...

Страница 7: ...ale 5 Zero the balance scale ZERO 6 Place reference sample quantity 20 on the balance scale 7 Select reference sample quantity in increments of 1 1 2 3 99 or F repeatedly briefly or in increments of 1...

Страница 8: ...mpty container on the balance scale 5 Tare the balance ZERO 6 Place the reference weight for 100 on the balance scale 7 If desired change the number of decimal places displayed 100 0 100 00 or 100 fac...

Страница 9: ...n the display application is active 4 If necessary zero the balance scale ZERO 5 Place oversized sample on balance scale 6 Start application program F Symbol flashes on the display the weight value is...

Страница 10: ...yed ENTER steadily symbol flashes 7 Remove sample from balance scale 8 Place the next sample on the balance scale in this example 575 g 9 Store value in memory Totalized stored weight ENTER is display...

Страница 11: ...ance scale ZERO 5 Start application program F 6 Confirm the display airval ENTER 7 Determine weight of the sample in air place sample on the balance scale 8 Store weight in air value ENTER 9 Remove sa...

Страница 12: ...different calibration weight value F repeatedly 5 Confirm calibration weight value and ENTER start calibration span adjustment After the zero point is stored the required calibration weight flashes on...

Страница 13: ...ons 1 2 Application filter 1 2 1 Final readout 1 2 2 Filling 1 3 Stability range 1 3 1 1 4 digit 1 3 2 1 2 digit 1 3 3 1 digit 1 3 4 2 digits 1 3 5 4 digits 1 5 Calib adjust linearization 1 5 1 Calibr...

Страница 14: ...play calErr Calibration parameter not met e g Calibrate only when zero is displayed Balance scale not zeroed Press ZERO to tare the balance scale Balance scale is loaded Unload the balance scale appEr...

Страница 15: ...of operation w 9V battery Alkaline approx h 11 11 14 Rechargeable fully ch NiMH avg h 2 5 2 5 4 Model VIC 5101 VIC 3101 VIC 1501 VIC 711 VIC 511 VIC 10KG VIC 6KG VIC 4KG Weighing capacity g 5100 3100...

Страница 16: ...request Accessories Options C Marking The balance scale complies with the following EC Directives and European Standards Council Directive 89 336 EEC Electromagnetic compatibility EMC Applicable Europ...

Отзывы: