257
8
Display Shows
Possible Cause(s)
Possible Solution(s)
Invalid Hyper
Values
The hyper warning limit value
must be greater than all of your
target ranges in the time block
settings.
• Reset the hyper warning limit or revise the target
ranges in time blocks and re-enter the hyper
warning limit.
• Enter a hyper warning limit that is above the
target ranges of your time blocks. See Chapter 5,
Changing Meter Settings.
Invalid Hypo Values The hypo warning limit value
must be less than all of your
target ranges in the time block
settings.
• Reset the hypo warning limit or revise the target
ranges in time blocks and re-enter the hypo
warning limit.
• Enter a hypo warning limit that is below the
target ranges of your time blocks. See Chapter 5,
Changing Meter Settings.
Invalid PIN
The wrong PIN was entered.
Select OK and re-enter the PIN shown on the pump
display.
Invalid Record
Time/Date
The time/date entered is invalid
(Add Data entries cannot be set
to occur in the future).
Re-enter time/date.
53895_00048010161_E.indb 257
9/4/15 9:34 AM
Содержание Aviva
Страница 6: ...iv 53895_00048010161_E indb 4 9 4 15 9 34 AM...
Страница 12: ...6 53895_00048010161_E indb 6 9 4 15 9 34 AM...
Страница 20: ...14 53895_00048010161_E indb 14 9 4 15 9 34 AM...
Страница 52: ...46 53895_00048010161_E indb 46 9 4 15 9 34 AM...
Страница 76: ...70 53895_00048010161_E indb 70 9 4 15 9 34 AM...
Страница 179: ...173 5 11 10 To return to the Settings screen press Press 53895_00048010161_E indb 173 9 4 15 9 34 AM...
Страница 268: ...262 53895_00048010161_E indb 262 9 4 15 9 34 AM...
Страница 282: ...276 53895_00048010161_E indb 276 9 4 15 9 34 AM...
Страница 305: ...299 V vibrate setting 199 W warnings list of 229 warranty 272 53895_00048010161_E indb 299 9 4 15 9 34 AM...
Страница 306: ...300 53895_00048010161_E indb 300 9 4 15 9 34 AM...