Display Shows
Possible Cause(s)
Possible Solution(s)
Communication
Warning
• Communication of bolus
data from the pump was not
successful. Therefore, pump
data are not available and
the bolus data may not be
accurate.
• Communication of bolus
data from the pump was not
successful. Therefore, pump
data are not available and
the active insulin amount
may not be accurate.
• Ensure the meter and pump have
Bluetooth
wireless technology turned on.
• Ensure the meter and pump are within
communication range.
• If the low battery icon is displayed on the meter,
the meter cannot communicate with the pump.
Replace the batteries.
Connection Lost
The connection was lost
between the meter and the
pump during the pairing process.
Therefore, the attempt to pair
the meter and pump was
unsuccessful.
• Restart the pairing process. See the Advanced
Owner’s Booklet.
• Ensure the meter and pump have
Bluetooth
wireless technology turned on.
• If the low battery icon is displayed on the meter,
the meter cannot communicate with the pump.
Replace the batteries.
Pump is out of range of the
meter.
Ensure the meter and pump are within
communication range.
53895_00048010161_E.indb 251
9/4/15 9:34 AM
Содержание Aviva
Страница 6: ...iv 53895_00048010161_E indb 4 9 4 15 9 34 AM...
Страница 12: ...6 53895_00048010161_E indb 6 9 4 15 9 34 AM...
Страница 20: ...14 53895_00048010161_E indb 14 9 4 15 9 34 AM...
Страница 52: ...46 53895_00048010161_E indb 46 9 4 15 9 34 AM...
Страница 76: ...70 53895_00048010161_E indb 70 9 4 15 9 34 AM...
Страница 179: ...173 5 11 10 To return to the Settings screen press Press 53895_00048010161_E indb 173 9 4 15 9 34 AM...
Страница 268: ...262 53895_00048010161_E indb 262 9 4 15 9 34 AM...
Страница 282: ...276 53895_00048010161_E indb 276 9 4 15 9 34 AM...
Страница 305: ...299 V vibrate setting 199 W warnings list of 229 warranty 272 53895_00048010161_E indb 299 9 4 15 9 34 AM...
Страница 306: ...300 53895_00048010161_E indb 300 9 4 15 9 34 AM...