252
Display Shows
Possible Cause(s)
Possible Solution(s)
Diary Entry Used
For Advice
The diary entry selected has
been used for bolus advice and
modifications are not allowed.
Carefully confirm all information involving bolus
advice. You cannot modify bolus advice related entries
on the meter.
Diary Results
Expired
Bolus advice is no longer
available for this result.
Review data that was saved in My Data. Then, decide
to modify data, add data, or begin a new bolus advice
session.
E-51
Bad Strip Error
• The test strip is damaged.
• The test strip is not properly
inserted into the meter.
If you see this error message before you applied
blood to the test strip:
• Remove the test strip and reinsert it, or replace it
if damaged. Check the use by date on the test
strip container. Do not use test strips past the
use by date. If the message reappears, contact
Roche.
53895_00048010161_E.indb 252
9/4/15 9:34 AM
Содержание Aviva
Страница 6: ...iv 53895_00048010161_E indb 4 9 4 15 9 34 AM...
Страница 12: ...6 53895_00048010161_E indb 6 9 4 15 9 34 AM...
Страница 20: ...14 53895_00048010161_E indb 14 9 4 15 9 34 AM...
Страница 52: ...46 53895_00048010161_E indb 46 9 4 15 9 34 AM...
Страница 76: ...70 53895_00048010161_E indb 70 9 4 15 9 34 AM...
Страница 179: ...173 5 11 10 To return to the Settings screen press Press 53895_00048010161_E indb 173 9 4 15 9 34 AM...
Страница 268: ...262 53895_00048010161_E indb 262 9 4 15 9 34 AM...
Страница 282: ...276 53895_00048010161_E indb 276 9 4 15 9 34 AM...
Страница 305: ...299 V vibrate setting 199 W warnings list of 229 warranty 272 53895_00048010161_E indb 299 9 4 15 9 34 AM...
Страница 306: ...300 53895_00048010161_E indb 300 9 4 15 9 34 AM...